FLAT PANEL DISPLAY (FPD) CONOSCOPE LENS

fpd conoscope
RADIANT VISION SYSTEMS FPD CONOSCOPE LENS

Fast, Precise, Low-In Cost View Angle Performance Measurement Solutions Offer For Display Applications

Radiant Vision Systems FPD Conoscope Lens is built and designed to measure the angular distribution of color, luminance and contrast from displays and display components. The conoscope lens captures a full cone of view angle data in a single measurement to ±70 degrees. This provides quick, accurate results, making FPD Conoscope Lens the system ideal for both R&D projects and in-line production quality control. The lens is well-suited to a wide range of display types including those based on LCD and OLED technologies, as well as backlights.

The FPD Conoscope Lens mounts directly to a Radiant Vision Systems ProMetric® Imaging Photometer or Colorimeterusing a custom bracket.  Mounting the lens directly to a ProMetric imaging photometer or colorimeter offers the most efficient means of measuring view angle performance using a turnkey, compact camera solution that requires minimal setup and integration time, equipment, or expertise. As part of the ProMetric family of imaging colorimeters and photometers, the FPD conoscope lens and ProMetric camera solution also comes with ProMetric Software, a powerful software suite that provides equipment control, data acquisition and image analysis. Learn more

APPLICATIONS
View angle performance measurement for LCD and OLED flat panel displays | Technology evaluation and characterization in display R&D and quality control | In-line quality control during display manufacturing

Radiant Vision FPD Conoscope Lens

  • Accurate, reliable measurement of luminance, chromaticity, and contrast to ±70° view angle
  • Faster than goniometric systems, capturing data for all angles simultaneously
  • Easily paired with ProMetric imaging colorimeters or photometers
  • Combination of low cost, high performance, and flexibility, designed for R&D and quality assurance labs
  • Easy-to-use measurement control and analysis software

Radiant Vision FPD Conoscope Lens

ParameterPM-CO-070-Y / PM-CO-070-I
Primary Application View angle measurement
Working Distance3 mm
Minimum sampling area11 mm diameter
View angle±70°
Resolution0.05° / CCD pixel
Luminance – Minimum 0.01 cd/m²
Luminance – Maximum3,000 cd/m² 
System Accuracy**0°~60° Inclination
Luminance (Y) ±4%
Color Coordinates (x,y) ±0.004
60°~70° Inclination
Luminance (Y) ±5%
Color Coordinates (x,y) ±0.007
Short-term RepeatabilityLuminance (Y) ±0.02%
Measurement Capabilities***Luminance, Radiance, Angular Contrast, CIE Chromaticity Coordinates, Correlated Color Temperature (CCT)

* Specifications subject to change without notice.
** Accuracy dependent on spectrum under test and color calibration methods applied
*** Color measurement available with I-series cameras only