Chroma Meter CL-200A (illuminance, chromaticity and color temperature)
A Compact, Lightweight Instrument With Detachable Receptor Head And Comes With Easy-To-Use Software
Konica Minolta CL-200A chroma meter is the successor to the CL-200.
The CL-200A compact body fits in your palm. It is battery operated so it can be taken along and used anyway. The CL-200A is designed to measure not only old typical light sources but also new light sources such as LED lighting and organic EL lighting too.
Konica Minolta CL-200A chroma meter measures illuminance (lux), chromaticity (xy), correlated color temperature (CCT), dominant wavelength (DW) and excitation purity (Pe) of various light sources. The CL-200A uses precision sensors coordinate with the International Commission of Illumination (CIE) define color matching functions, the results are displayed in a variety of application specific color notation.
APPLICATIONS:
Light Emitting Diode (LED) billboard | Organic LED (OLED) | Electroluminescence (EL) lighting | Building and interior lighting research | Projector light source research and color inspection | Color viewing cabinet maintenance | Evaluating illuminance distribution of lighting fixtures | Evaluating light distribution characteristics of LED illumination module

CL-200A Chroma Meter
- Compact – hand-held, lightweight design
- Portable – battery operated, receptor head can also be detached from the main body for remote measurements
- User-friendly – low battery indicator, can be controlled by a PC, easily transfer data to Excel, and includes a calibration function
- Simultaneous measurements – has the ability to determine illuminance, color temperature, chromaticity, dominant wavelength, and excitation purity of a light source
- Ideal for white LEDs – can measure the chromaticity from the phosphor as well as inspect the final product
- Multi-use – can be used for any light source, ideal for LEDs and OLEDs
- LED ranking function – using the included software, users can rank measurements on a chromaticity chart for easy comparison
CL-200A Chroma Meter
Relative Spectral Response | Closely matches CIE Standard Observer curves X(λ), y(λ), and Z(λ) Within 6%(f1’) of the CIE spectral luminous effeciency V(λ) |
Cosine response(f2) | Ev: Within 3% |
Receptor | Silicon photocell |
Measuring function | Tristimulus values:XYZ Chromaticity : Ev xy, Ev u’v’ Correlated color temperature: Ev, Tcp, Δuv Color difference: Δ(XYZ), Δ(Ev, xy), Δ(Ev u’v’), ΔEv Δu’v’ |
Other function | User calibration function, Data hold function, Multi-point measurement(2 to 30 points) |
Measuring range | 0.1∼99,990lx, 0.01∼9,999 fcd(Chromaticity:5 lx, 0.5 fcd or above) in four automatically selected ranges(lx or fcd is swichable) |
Accuracy | Ev: ±2%±1digit of displayed value xy: ±0.002(800lx, standardilluminant A measured) |
Repeatability | Ev: 0.5%+1 digit(2σ) xy: ±0.0005 |
Temperature drift | Ev: ±3%±1digit of displayed value, xy: ±0.003 |
Humidity drift | Ev: ±3%±1digit of displayed value, xy: ±0.003 |
Response time | 0.5 seconds (each measurement) |
Digital output | USB 2.0 full-speed |
Display | 4 Significant-digit LCD with back-light illumination |
Operating temperature/ Humidity range | -10 to 40°C, relative humidity 85% or less (at 35°C) with no condensation |
Storage temperature/ Humidity range | -20 to 55°C, relative humidity 85% or less (at 35°C) with no condensation |
Power source | 2 AA-size batteries/AC adapter (optional) |
Battery life | 72 hours or longer (when alkaline bateries are used) in continuous measurement |
Dimensions | 69 x 174 x 35 mm |
Weight | 215g(7.6 oz.) not including batteries |
Standard accessories | Case T-A10, Cap T-A13, Neck Strap, batteries, Data Management Software CL-S10w, Connecting Cable T-A15 |
Optional accessories | Recepter Head, Adapter Unit for Main Body T-A20, Adapter Unit for Receptor Head T-A21, AC Adapter AC-308(available for 1 to 10 Receptor Heads) AC Adapter AC-311(available for 1 to 30 Receptor Heads) Printer Cable T-A12, Hood CL-A11, Case CL-A10 |
Data Management Software CL-S107w Specification
Software type | MicroSoft Excel add-in software Excel is required for operation of this software |
Operating Systems | Windows XP + Excel 2003 (Japanese, English,Chinese) Windows 7 + Excel 2003 (Japanese, English, Chinese) |
Hardware to connect | CL-200A CL-200 some of the functions are not available when connected with the CL-200 |
∗Specfications are subject to change without prior notice.